HP PRO Solution Put to the Test

Stable and reliable immobilization for proton and photon therapy

Reliability and stability – two pillars of safe and effective immobilization for proton/photon therapy. Characteristics that were of the highest importance to us when we designed the Orfit HP PRO – our proton compatible immobilization solution.

A series of extensive clinical tests by MAASTRO Clinic (Maastricht, The Netherlands), has now further confirmed the reliability and stability of the HP PRO.

 

Proton compatible immobilization

The Orfit HP PRO Solution is an immobilization device designed for brain, head, and neck proton therapy. As it is a versatile system, it lends itself perfectly to photon therapy as well.

The system has been designed to meet the specific standards that proton therapy demands. It includes a homogeneous low-density immobilization device with a narrow carbon fiber base plate in the head and neck area. Because of its ingenious design, it allows for close range positions of the treatment device.

Orfit HP PRO, a proton compatible immobilization system

In addition, the HP PRO System uses the ultra-stable Nanor masks for guaranteed comfort and stability. This innovative thermoplastic material makes use of nanotechnology for improved mechanical properties.

Learn more about the Orfit HP PRO system on orfit.com/proton.

 

Clinical testing

MAASTRO Clinic recently did extensive testing on the Orfit HP PRO as a setup for proton head and neck treatments.

The clinic has been using the Orfit HP PRO system since September 2018 and they apply it for both photon and proton therapy.

In their internal clinical test, MAASTRO verified the efficacy of two different Orfit HP PRO set-ups:

 

HP PRO head immobilization

The first set-up consisted of the following components:

First proton compatible immobilization set-up put to the test.

 

HP PRO head, neck and shoulders immobilization

The second set-up consisted of the following components:

Second proton compatible immobilization set-up put to the test.

 

Method and Results

In both set-up analyses, MAASTRO used extra CBCT imaging before and after treatment to identify residual error and intrafraction movement.

In the head and neck set-up, they applied 6D offline matches of 23 patients in total to determine the margins in lateral (x), longitudinal (y), and vertical (z) direction. They found that a PTV margin of 1 mm is achievable to treat neuro patients with the HP PRO immobilization system, extended with a biteblock.

In the head, neck and shoulders set-up, they applied 4D offline matches of 17 patients to determine the margins in lateral (x), longitudinal (y), and vertical (z) direction. For this set-up, MAASTRO found that a margin of 3 mm CTV-PTV is suitable to treat the H&N patient.

 

In conclusion, MAASTRO’s findings suggest that the HP PRO system is indeed a reliable and stable system, resulting in small margins for CTV-PTV.

>> Review MAASTRO’s complete findings on the HP PRO’s efficacy in these slides:

(click to enlarge)

HP PRO for proton therapy put to the test

 

 

Written by Rob Martens, Bachelor in Radiation Oncology and Medical Imaging Techniques

Radiation Oncology Product Specialist at Orfit Industries.

Rob has over 10 years of experience as a Radiation Therapist in a clinical and research environment.

He provides product education in Dutch Radiation Oncology Centres and gives input and support to the Orfit R&D and Sales and Marketing team.

 

We've selected some similar articles for you
The AIO Solution 3.0 Allows for a Better Workflow at Aurora BayCare
An interview with Anna Schneider, Radiation Therapist II, at Aurora BayCare Medical Center.
October 25, 2024 – We sat down with Anna Schneider, a radiation therapist at Aurora BayCare in Green Bay, Wisconsin. Anna has been a therapist for about 11 years now and has been using Orfit solutions at Aurora BayCare since about 2017. The center has 2 Varian-Truebeam linear accelerators with…
Read more
How Orfit Solutions Help Both Patients and Radiation Therapists
An interview with Jennifer Mundine and Veronica Shilling, radiation therapists at CHI St. Joseph Health Bryan Hospital.
Norfolk, VA – September 5, 2024 – We sat down with Jennifer Mundine and Veronica Shilling, radiation therapists at St. Joseph Health Bryan Hospital at College Station, Texas. Jennifer and Veronica have 13 years, and more than 20 years of experience, respectively. Both RTTs use Orfit solutions every day. College Station is…
Read more
Indian Hospital & Research Centre Enhances Patient Care with Orfit Products
Strategies for Treating 70+ Patients Daily on a Single LINAC.
High Volume, High Efficiency Wijnegem, June 11, 2024 – Prakash Umbarkar, a senior radiation therapist at Mumbai’s P.D. Hinduja National Hospital provides his perspective on the company’s immobilization and positioning solutions. Part of the Hinduja Group, P. D. Hinduja Hospital & Medical Research Centre is topmost ranked among the leading…
Read more